Abstract
This paper describes design and optimization for the wide angle spin-echo spectrometer (WASP) presently being planned at the ILL. The spectrometer will be a replacement for the high intensity spectrometer IN11 and will enhance its signal by more than one order of magnitude while maintaining the resolution of the present spectrometer. The paper outlines the magnetic field calculations and the considerations about possible limitations. The projected performance has been verified carefully by Monte Carlo raytracing and Biot-Savart magnetic field calculations. The maximum momentum transfer of the new spectrometer is to be extended to 4Å-1.
Original language | English |
---|---|
Pages (from-to) | 39-47 |
Number of pages | 9 |
Journal | Journal of Neutron Research |
Volume | 15 |
Issue number | 1 |
DOIs | |
State | Published - 2007 |
Externally published | Yes |
Keywords
- Instrumentation
- Magnetic field calculations
- Neutron scattering
- Neutron spin-echo