Abstract

This paper reports the detailed noise characterization, investigation of various noise sources, and its mitigation to improve the performance of a cryogenic bolometer detector for the TIN.TIN experiment. The noise spectrum has been measured for a sapphire bolometer test setup with an indigenously developed neutron transmutation doped Ge sensor in the cryogen free dilution refrigerator system at Mumbai. The effect of external noise, arising either from ground loops in the system or from the diagnostic and control electronics of the cryostat, on the performance of the bolometer is assessed. A systematic comparison of the influence of different noise pickups on the bolometer resolution is also presented. The best-achieved resolution (σE) at 15 mK is ∼15 keV for heater pulses and appears to be mainly limited by the noise due to the pulse tube cryocooler.

Original languageEnglish
Article number096104
JournalReview of Scientific Instruments
Volume90
Issue number9
DOIs
StatePublished - Sep 1 2019
Externally publishedYes

Fingerprint

Dive into the research topics of 'Study of the effect of external noise pickups on the performance of a cryogenic bolometer'. Together they form a unique fingerprint.

Cite this