Spatial transfer function for the beam emission spectroscopy diagnostic on DIII-D

M. W. Shafer, R. J. Fonck, G. R. McKee, D. J. Schlossberg

Research output: Contribution to journalArticlepeer-review

35 Scopus citations

Abstract

The spatial transfer function for the beam emission spectroscopy (BES) diagnostic is critical to quantitatively interpret local density fluctuation measurements. A three-dimensional geometrical calculation of the spatial transfer function is presented for the upgraded BES diagnostic on DIII-D to determine its spatial resolution and wave-number sensitivity. The spatial transfer function calculation for the BES system on DIII-D incorporates the high speed (f2) collection optics, an optical fiber bundle, neutral beam-sight line geometry, the neutral beam cross-section intensity profile, magnetic field pitch angle, as well as atomic physics of the finite atomic transition time of the collisionally excited beam atoms. The resulting imaged volumes for each BES channel typically have ∼1-2 cm radial and poloidal resolutions. In addition, the viewing volume is nominally aligned along a magnetic field line to minimize spatial smearing of the field-aligned turbulent eddies. This calculation is crucial for performing the proper deconvolution of measured signals to determine correlation lengths and wave-number spectra of the measured density fluctuations, and is also required for applying synthetic diagnostics to nonlinear turbulence simulation codes (e.g., GYRO) to relate code output to measured signals.

Original languageEnglish
Article number10F110
JournalReview of Scientific Instruments
Volume77
Issue number10
DOIs
StatePublished - 2006
Externally publishedYes

Funding

This work was supported by U.S. DOE under DE-FG03-96ER54373.

FundersFunder number
U.S. Department of EnergyDE-FG03-96ER54373

    Fingerprint

    Dive into the research topics of 'Spatial transfer function for the beam emission spectroscopy diagnostic on DIII-D'. Together they form a unique fingerprint.

    Cite this