Abstract
We have experimented with a radio-frequency-based streak camera. The basic idea behind the method consists of magnetic deflection of the electron beam imparted by a radio-frequency cavity. All the components of the camera were tested and the time resolution was measured to lie within the sub-ps domain. Further advantages of the method are simplification of triggering, weaker dependence on space-charge effect within the electron beam, high modularity, and ease of implementation.
| Original language | English |
|---|---|
| Pages (from-to) | 2622-2626 |
| Number of pages | 5 |
| Journal | Review of Scientific Instruments |
| Volume | 70 |
| Issue number | 6 |
| DOIs | |
| State | Published - Jun 1999 |
| Externally published | Yes |