Scatter limitation to correct for arm movement in PET/CT

Jason Callahan, David Binns, Timothy Deller, Rodney J. Hicks

Research output: Contribution to journalArticlepeer-review

4 Scopus citations

Abstract

A commonly seen artifact in PET/CT scanning is areas of "washed out" activity on the reconstructed PET scan due to arm movement after the CT scan. When a patient moves his/her arms after the CT scan, there are areas of activity outside the mask defined by the CT. This leads to an overestimation of scatter in these areas creating significant reconstruction artifacts. This case presents an example of a novel processing algorithm known as "scatter limitation" that corrects the overestimation of scatter caused by arm movement.

Original languageEnglish
Pages (from-to)786-787
Number of pages2
JournalClinical Nuclear Medicine
Volume37
Issue number8
DOIs
StatePublished - Aug 2012
Externally publishedYes

Keywords

  • PET artifact
  • arm movement
  • scatter limitation