Reconstruction of multi-energy x-ray computed tomography images of laboratory mice

    Research output: Contribution to journalArticlepeer-review

    34 Scopus citations

    Abstract

    A new x-ray computed tomography (CT) system is being developed at Oak Ridge National Laboratory to image laboratory mice for the purpose of rapid phenotype screening and identification. One implementation of this CT system allows simultaneous capture of several sets of sinogram data, each having a unique x-ray energy distribution. The goals of this paper are to (1) identify issues associated with the reconstruction of this energy-dependent data and (2) suggest preliminary approaches to address these issues. Due to varying numbers of photon counts within each set, both traditional (filtered backprojection, or FfiR) and statistical (maximum likelihood, or Ml. ) tomographic image reconstruction techniques have been applied to the energy-dependent sinogram data. Results of reconstructed images using both algorithms on sinogram data (high- and low-count) are presented. Also, tissue contrast within the energy-dependent images is compared to known x-ray attenuation coefficients of soft tissue (e. g. muscle, bone, and fat).

    Original languageEnglish
    Pages (from-to)1081-1086
    Number of pages6
    JournalIEEE Transactions on Nuclear Science
    Volume46
    Issue number4 PART 2
    DOIs
    StatePublished - 1999

    Funding

    'Research sponsored by the ILaboratory Dirccted Iccsearch and Ilevelopmenl I'rogram or Oak Ridge National Laboratory, managed by Lockhced Martin Energy Research Corp. for the U.S. Department of Energy iindcr Contract No. UI~~,-AC05-960R22464.

    Fingerprint

    Dive into the research topics of 'Reconstruction of multi-energy x-ray computed tomography images of laboratory mice'. Together they form a unique fingerprint.

    Cite this