@inproceedings{79d63b958e9941518b11427ae18e624a,
title = "New X-ray computed tomography system for laboratory mouse imaging",
abstract = "Two versions of a new high-resolution x-ray computed tomography system are being developed to screen mutagenized mice in the Oak Ridge National Laboratory Mouse Genetics Research Facility. The first prototype employs a single pixel solid-state detector with a pinhole collimator operating in pulse counting mode. The second version employs a phosphor screen/CCD detector operating in current mode. The gantry consists of a low-energy x-ray tube, two linear translation stages and a rotational stage. For the single pixel detector image resolution is determined by the step size of the detector stage; preliminary images have been acquired at 100 μm and 250 μm resolutions. The resolution of the phosphor screen detector is determined by the modulation transfer function of the phosphor screen; images with resolutions approaching 50 μm have been acquired. The system performance with the two detectors is described and recent images are presented.",
author = "Paulus, {M. J.} and H. Sari-Sarraf and Gleason, {S. S.} and M. Bobrek and Hicks, {J. S.} and Johnson, {D. K.} and Behel, {J. K.} and Thompson, {L. H.}",
year = "1999",
language = "English",
isbn = "0780350227",
series = "IEEE Nuclear Science Symposium and Medical Imaging Conference",
publisher = "IEEE",
pages = "1259--1263",
booktitle = "IEEE Nuclear Science Symposium and Medical Imaging Conference",
note = "Proceedings of the 1998 IEEE Nuclear Science Symposium Conference Record ; Conference date: 08-11-1998 Through 14-11-1998",
}