Abstract
Computer Assisted Detection Systems (CAD) in mammography incorporate two critical stages: (i) prescreening to localize suspicious regions and (ii) detailed analysis of the regions for false positive reduction. In this work, we present a new technique for automatic detection of suspicious masses for prescreening mammograms. The hypothesis of the proposed technique is that malignant masses manifestate as superimposed concentric layers. Morphological characterization of these layers can form the foundation of an automated scheme for detection of suspicious masses. The study was based on fifty nine screening mammograms from the Digital Database of Screening Mammography (DDSM). Overall, the proposed scheme performs at 85.7% sensitivity with an average of 0.53 false positives per image. The scheme targets malignant masses and , thus it can provide a second opinion to radiologists without sending benign masses to unnecessary biopsy.
Original language | English |
---|---|
Pages (from-to) | 1279-1282 |
Number of pages | 4 |
Journal | Conference proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Conference |
Volume | 26 II |
State | Published - 2004 |
Externally published | Yes |
Event | Conference Proceedings - 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2004 - San Francisco, CA, United States Duration: Sep 1 2004 → Sep 5 2004 |
Keywords
- Breast Cancer
- Computer Assisted Detection
- Masses
- Morphological Concentric Layer Analysis