Measurements of scatter peaks in137Cs and60Co sources

    Research output: Contribution to journalArticlepeer-review

    3 Scopus citations

    Abstract

    Results from tests of radiation detection instruments with radionuclide identification capabilities will depend on the sources used for the tests. Radionuclide identification detectors are designed to measure photons and provide an identification of the source being measured. High-resolution spectra need to be acquired to determine all the observable peaks in the source spectra before testing these types of instruments. These peaks may be due to impurities and/or scatter peaks in the sources. This paper discusses the issues encountered with the response of a radioisotope identification device due to scatter peaks in one type of source used for testing. In addition, it provides spectra for different source types and source constructions to compare the differences in scatter, allowing for a better source type selection for instrument testing.

    Original languageEnglish
    Article number123012
    JournalJournal of Research of the National Institute of Standards and Technology
    Volume123
    DOIs
    StatePublished - Jun 28 2018

    Keywords

    • ANSI N42
    • Radiation detection for homeland security
    • Scatter peaks
    • Source spectra
    • Standards
    • Testing of radiation detection equipment