Measurements of scatter peaks in137Cs and60Co sources

Leticia Pibida, David West

Research output: Contribution to journalArticlepeer-review

3 Scopus citations

Abstract

Results from tests of radiation detection instruments with radionuclide identification capabilities will depend on the sources used for the tests. Radionuclide identification detectors are designed to measure photons and provide an identification of the source being measured. High-resolution spectra need to be acquired to determine all the observable peaks in the source spectra before testing these types of instruments. These peaks may be due to impurities and/or scatter peaks in the sources. This paper discusses the issues encountered with the response of a radioisotope identification device due to scatter peaks in one type of source used for testing. In addition, it provides spectra for different source types and source constructions to compare the differences in scatter, allowing for a better source type selection for instrument testing.

Original languageEnglish
Article number123012
JournalJournal of Research of the National Institute of Standards and Technology
Volume123
DOIs
StatePublished - Jun 28 2018

Keywords

  • ANSI N42
  • Radiation detection for homeland security
  • Scatter peaks
  • Source spectra
  • Standards
  • Testing of radiation detection equipment

Fingerprint

Dive into the research topics of 'Measurements of scatter peaks in137Cs and60Co sources'. Together they form a unique fingerprint.

Cite this