Measurement of neutral gas pressure in the D-module of GAMMA 10/PDX by using ASDEX type fast ionization gauge

K. Ichimura, M. Fukumoto, M. M. Islam, M. S. Islam, K. Shimizu, K. Fukui, M. Ohuchi, K. Nojiri, A. Terakado, M. Yoshikawa, N. Ezumi, M. Sakamoto, Y. Nakashima

Research output: Contribution to journalArticlepeer-review

9 Scopus citations

Abstract

In the divertor simulation experiments in the GAMMA 10/PDX tandem mirror, pressure of the neutral gas was investigated by using a fast ionization gauge. The gauge was absolutely calibrated for hydrogen gas by using a capacitance manometer. Change of the gauge sensitivity due to the magnetic field of GAMMA 10/PDX was also evaluated. The typical gas pressure measured in detached plasma experiments was 0.1-10 Pa. The degree of plasma detachment determined from the reduction of heat flux was enhanced as the gas pressure increases. Rapid increase of the gas pressure under the plasma flow was also observed.

Original languageEnglish
Article number11D424
JournalReview of Scientific Instruments
Volume87
Issue number11
DOIs
StatePublished - Nov 1 2016
Externally publishedYes

Fingerprint

Dive into the research topics of 'Measurement of neutral gas pressure in the D-module of GAMMA 10/PDX by using ASDEX type fast ionization gauge'. Together they form a unique fingerprint.

Cite this