Abstract
Large area avalanche photodiode detector array upgrade for a ruby-laser Thomson scattering (TS) system to increase spectral coverage and substantially improve the signal-to-noise ratio (SNR) was discussed. The spectral resolution was increased from five channels on the long-wavelength side to an 11-channel system. Scattered light collection was also upgraded, allowing the radial viewing location of the TS system to be easily changed between plasma discharges.
| Original language | English |
|---|---|
| Pages (from-to) | 1649-1652 |
| Number of pages | 4 |
| Journal | Review of Scientific Instruments |
| Volume | 74 |
| Issue number | 3 II |
| DOIs | |
| State | Published - Mar 2003 |
| Externally published | Yes |