Abstract
We have made feasibility study of anomalous small-angle X-ray scattering (ASAXS) near the sulfur K-edge (2.472 keV). In the present study, we report the improvement of SAXS measurement with the use of Si (111) channel-cut monochromator and Kratky camera at BL27SU, SPring-8. The photon flux and the energy resolution of the incident X-ray beam increased by 10 times and 2.5 times, respectively. SAXS intensity also increased by about 10 times with the improved low-q and high-q limits, enabling us to measure SAXS with a better S/N ratio over a wider q range.
Original language | English |
---|---|
Article number | 012014 |
Journal | Journal of Physics: Conference Series |
Volume | 272 |
Issue number | 1 |
DOIs | |
State | Published - 2011 |