Improved lipid analysis using a 2D-LC-MS system with a novel injection procedure

József Pánczél, Vilmos Kertesz, Matthias Schiell

Research output: Contribution to journalArticlepeer-review

Abstract

The aim of this study was to improve analysis of nonpolar lipidomics sample extracts using reversed phase (RP) chromatography. A 4/3/3 (v/v/v) mixture of methanol/methyl tert-butyl ether/chloroform (MeOH/MTBE/CHCl3, MMC) was chosen for sample extraction solvent based on its proven extraction capability for several lipid classes. To avoid carry over, loss of analytes and peak distortion the loops and all capillaries of the presented LC system were flushed and filled up with methanol until the analytical column. The choice of methanol was due to its weak elution strength and being infinitely miscible with MMC and several other nonpolar solvents. This allowed injection of a 100 μl sample that was 20 μl nonpolar extraction solvent diluted fivefold with methanol. All lipids of 25 lipid classes were transferred quantitatively to the column head where the online dilution of methanol was carried out with aqueous eluent for focusing the lipid analytes. The weak elution strength of methanol prevented peak distortions. The consecutive reversed phase elution resulted in remarkably narrow peaks (full width at half maximum was 0.07–0.08 min typically) and enhanced sensitivity (limit of detection usually in sub nM region) because of increased sample injection volume and narrow peaks. Calibration and quality control samples made by diluting commercial lipid standards 200–50000 times confirmed the applicability of this approach both for targeted lipid quantification and for untargeted quantitative comparison of lipids from different sources.

Original languageEnglish
Article number124129
JournalJournal of Chromatography B: Analytical Technologies in the Biomedical and Life Sciences
Volume1239
DOIs
StatePublished - May 15 2024

Keywords

  • Lipidomics
  • Mass spectrometry
  • Reversed phase (RP) chromatography
  • Signal enhancement

Fingerprint

Dive into the research topics of 'Improved lipid analysis using a 2D-LC-MS system with a novel injection procedure'. Together they form a unique fingerprint.

Cite this