Abstract
High resolution spectroscopy on the Tokamak à Configuration Variable (TCV) divertor plasma provided Doppler broadening measurements to infer the ion and neutral temperature of injected helium gas. This paper presents the Divertor Spectroscopy System's (DSS) access to He II ion temperature measurements over a broad range, ≈0.5-15 eV, with an uncertainty of <10% for most of the studied plasma discharges. TCV's shaping flexibility was employed to validate these measurements against Thomson scattering across the DSS lines of sight. In detachment-related experiments, Ti(He II) ≃ Te, making this diagnostic a reliable thermometer along the divertor leg plasma over the wide range of magnetic equilibria and divertor configurations achievable in TCV. A detailed description of the diagnostic hardware, data analysis, and sources of uncertainty is presented.
Original language | English |
---|---|
Article number | 123505 |
Journal | Review of Scientific Instruments |
Volume | 93 |
Issue number | 12 |
DOIs | |
State | Published - Dec 1 2022 |
Funding
This work was carried out within the framework of the EUROfusion Consortium, funded by the European Union via the Euratom Research and Training Program (Grant Agreement No. 101052200—EUROfusion). Views and opinions expressed are however those of the author(s) only and do not necessarily reflect those of the European Union or the European Commission. Neither the European Union nor the European Commission can be held responsible for them. This work was supported, in part, by the Swiss National Science Foundation.