Estimability index for volume quantification of homogeneous spherical lesions in computed tomography

Ehsan Samei, Marthony Robins, Baiyu Chen, Greeshma Agasthya

Research output: Contribution to journalArticlepeer-review

4 Scopus citations

Abstract

Volume of lung nodules is an important biomarker, quantifiable from computed tomography (CT) images. The usefulness of volume quantification, however, depends on the precision of quantification. Experimental assessment of precision is time consuming. A mathematical estimability model was used to assess the quantification precision of CT nodule volumetry in terms of an index (e′), incorporating image noise and resolution, nodule properties, and segmentation software. The noise and resolution were characterized in terms of noise power spectrum and task transfer function. The nodule properties and segmentation algorithm were modeled in terms of a task function and a template function, respectively. The e′ values were benchmarked against experimentally acquired precision values from an anthropomorphic chest phantom across 54 acquisition protocols, 2 nodule sizes, and 2 volume segmentation softwares. e′ exhibited correlation with experimental precision across nodule sizes and acquisition protocols but dependence on segmentation software. Compared to the assessment of empirical precision, which required ∼300 h to perform the segmentation, the e′ method required ∼3 h from data collection to mathematical computation. A mathematical modeling of volume quantification provides efficient prediction of quantitative performance. It establishes a method to verify quantitative compliance and to optimize clinical protocols for chest CT volumetry.

Original languageEnglish
Article number031404
JournalJournal of Medical Imaging
Volume5
Issue number3
DOIs
StatePublished - Jul 1 2018
Externally publishedYes

Funding

The authors have no direct conflict of interest. E.S. has research funding from GE Healthcare and Siemens Heathineers unrelated to the topic of this study.

FundersFunder number
GE Healthcare and Siemens Heathineers

    Keywords

    • biomarker
    • computed tomography nodule volume quantification
    • estimability index (e ′)
    • noise power spectrum
    • precision
    • quantitative imaging volumetry
    • task transfer function

    Fingerprint

    Dive into the research topics of 'Estimability index for volume quantification of homogeneous spherical lesions in computed tomography'. Together they form a unique fingerprint.