Microchannel plate photodetectors, capable of picosecond time resolution and sub-mm spatial resolution, are a perfect candidate for the next generation of photodetectors for precision timing measurements. Argonne National Laboratory is producing low-cost, all-glass body, planar photodetectors with an indium seal. The design and fabrication of 6-cm square photodetectors has been well demonstrated in an ultra-high vacuum system. Recently, a new design was developed to optimize the photodetector bias voltage. This design offers an improved configuration and allows external control of the bias voltage for each internal detector component. Design and measurements of the new independently biased devices are described in this paper. We performed a systematic study on the bias voltage and achieved a gain on the order of 10^{{\mathbf {7}}} , a time resolution better than 35 ps, and a spatial resolution better than 1 mm.

Original languageEnglish
Article number7756301
Pages (from-to)1871-1879
Number of pages9
JournalIEEE Transactions on Nuclear Science
Volume64
Issue number7
DOIs
StatePublished - Jul 2017
Externally publishedYes

Funding

Manuscript received July 1, 2016; revised October 7, 2016; accepted November 9, 2016. Date of publication November 24, 2016; date of current version July 14, 2017. This work was supported by the U.S. Department of Energy, Office of Science, Office of Basic Energy Sciences and Office of High Energy Physics under Contract DE-AC02-06CH11357. (Corresponding author: Jingbo Wang.) The authors are with High Energy Physics division, Argonne National Laboratory, 9700 South Cass Avenue, Lemont, IL 60439 USA (e-mail: [email protected]; [email protected]; [email protected]; rdharmapalan@ anl.gov; [email protected]; [email protected]; [email protected]; rgwcdf@ hep.anl.gov; [email protected]; [email protected]; [email protected]; [email protected]).

FundersFunder number
Office of Basic Energy Sciences
U.S. Department of Energy
Office of Science
High Energy PhysicsDE-AC02-06CH11357

    Keywords

    • Bias voltage
    • gain
    • microchannel plate
    • optimization
    • photodetector
    • spatial resolution
    • time resolution

    Fingerprint

    Dive into the research topics of 'Design Improvement and Bias Voltage Optimization of Glass-Body Microchannel Plate Picosecond Photodetector'. Together they form a unique fingerprint.

    Cite this