Characterizing the timing performance of a fast 4H-SiC detector with an 241Am source

Xiaodong Zhang, Joshua W. Cates, Jason P. Hayward, Giuseppe Bertuccio, Donatella Puglisi, Paul A. Hausladen

Research output: Contribution to journalArticlepeer-review

18 Scopus citations

Abstract

An SPX4 4H-silicon carbide detector consisting of 4 × 4 pixels was developed and studied experimentally. Its pixel size is 400 × 400μm2. A timing resolution of 117 ± 11 ps full width at half-maximum (FWHM) has been measured for the detection of alphas. With such good timing performance and high granularity, the SiC pixel detector holds great promise as an associated alpha-particle detector for fast neutron imaging.

Original languageEnglish
Article number6520962
Pages (from-to)2352-2356
Number of pages5
JournalIEEE Transactions on Nuclear Science
Volume60
Issue number3
DOIs
StatePublished - 2013
Externally publishedYes

Keywords

  • Associated particle imaging
  • neutron radiography
  • semiconductor radiation detectors
  • silicon carbide
  • timing resolution

Fingerprint

Dive into the research topics of 'Characterizing the timing performance of a fast 4H-SiC detector with an 241Am source'. Together they form a unique fingerprint.

Cite this