Abstract
A bright neutron source such as the Spallation Neutron Source (SNS) places extreme requirements on detectors including excellent 2-D spatial imaging and high dynamic range. Present imaging detectors have either shown position resolutions that are less than acceptable or they exhibit excessive paralyzing dead times due to the brightness of the source. High neutron detection efficiency with good neutron-gamma discrimination is critical for applications in neutron scattering research where the usefulness of the data is highly dependent on the statistical uncertainty associated with each detector pixel. A detector concept known as MicroMegas (MicroMEsh GAseous Structure) has been developed at CERN in Geneva for high-energy physics charged-particle tracking applications and has shown great promise for handling high data rates with a rather low-cost structure. We are attempting to optimize the MicroMegas detector concept for thermal neutrons and have designed a 1-D neutron strip detector which we have tested. In addition, we are performing research into the compatibility of various converter coatings. Our goal is to develop a manufacturable detector that could be scaled to a 1m2, 2-D array for use at the SNS and other facilities.
Original language | English |
---|---|
Article number | N17-2 |
Pages (from-to) | 360-364 |
Number of pages | 5 |
Journal | IEEE Nuclear Science Symposium Conference Record |
Volume | 1 |
DOIs | |
State | Published - 2003 |
Event | 2003 IEEE Nuclear Science Symposium Conference Record - Nuclear Science Symposium, Medical Imaging Conference - Portland, OR, United States Duration: Oct 19 2003 → Oct 25 2003 |