基于Davies-Gray电位滴定法精密测定25~50 mg铀含量

Translated title of the contribution: Precision Measuring 25-50 mg Uranium Concentration by Davies-Gray Potentiometric Titration Method

Lan Wang, Chun Li, Ji Long Zhang, Rogers Kayron, Li Fang Yang, Zhao Hui Wu, Li Kun Liu

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

Uranium is an important material of nuclear fuel. Precise and accurate measurement of uranium concentration is critical to improving the nuclear facilities' capabilities to effectively fulfill their responsibility of nuclear safeguard. The Davies-Gray potentiometric titration method is optimized by modifying the stirring speed and period and replacing the commonly used composite electrode with pure platinum wire and then the measurement method of 25-50 mg uranium sample is established. When this method is used for titrating 25-50 mg uranium solution, the relative standard deviation is 0.056 8%(n=12), which is more precise than any other methods available at home for gram level uranium titration. Besides, this method is also used for measuring the uranium concentration in UO2 pellets during inter-laboratory capability comparisons hold by IAEA in 2017. The Z-score is 0.78, less than 2, which is satisfactory.

Translated title of the contributionPrecision Measuring 25-50 mg Uranium Concentration by Davies-Gray Potentiometric Titration Method
Original languageChinese (Traditional)
Pages (from-to)573-577
Number of pages5
JournalHe-Huaxue yu Fangshe Huaxue/Journal of Nuclear and Radiochemistry
Volume41
Issue number6
DOIs
StatePublished - Dec 20 2019

Keywords

  • 25-50 mg uranium concentration
  • Davies-Gray potentiom-etric titration
  • Experimental study

Fingerprint

Dive into the research topics of 'Precision Measuring 25-50 mg Uranium Concentration by Davies-Gray Potentiometric Titration Method'. Together they form a unique fingerprint.

Cite this